Housing | Stainless steel 1.4404 (316L), Aluminum, Type 6P/ IP66/ IP68 |
Certificates | CE, ATEX, IEC-Ex, FM |
Process Temperature Range | -20°C to +250°C |
Pressure Range | -1 .. +100bar max. |
Supply Voltage | 9.6 … 35 V DC, 2-wire loop |
Measuring Range | Rope 60m |
Material Probe | SS 316 |
Material Sealing | FFKM |
Second line of defense* (optional) | Borosilicate glass GPC 540 with 316L |
Material Isolation | Peek |
Signal Output | 4 .. 20mA, 2-Wire |
Communication | HART |
Sensitivity | DK ≥1,6 |
Process Connection | ≥ NPT ¾", ≥ G ¾", different flanges |
Programming | Via display and programming buttons or via HART interface and FDT software. |
• Rope version.
• Works in applications with vapour, buildup, foam generation and condensation.
• Easy setting and diagnosis with control module and display.
• Process temperature up to 250°C.
• Process pressure up to 100bar.
• A variety of approvals available.
• High chemical resistance of the probe.