DIMF 1.3

Sold By :

rjc.co.th



The liquid density sensor of the DIMF series is used for continuous measurement of the density / concentration of liquids or liquid mixtures. The proven tuning fork principle guarantees a high measuring accuracy with very good long-term stability. The robust design ensures reliable operation even under harsh operating conditions.


Nominal size10mm
Process connection

Internal thread G ¼ ISO 228

Flanges DN10 according to pressure stage PN 40 (or Class 150/300 RF) according to DIN 2501 (or ANSI B 16.5)

Other pressure ratings available on request

MaterialSpecial alloy NiFeCr and 1.4571 / 316 Ti
Density range0 to 5000 kg/m³ (0 Ib/ft³ to 11023.11 Ib/ft³)
Calibration range400 to 2000 kg/m³ (881.85 Ib/ft³ to 4409.25 Ib/ft³)
Accuracyup to ±0.01% (±0.1kg/m³ – ±0,22 Ib/ft³)
Repeatabilityup to ±0.005% (±0.05kg/m³ – ±0.11 Ib/ft³)
Process temperature-40°C to +150°C (-40°F to + 302)
Process pressure100 bar (1450 psi)
OutputWith transmitter electronics:
2-wire current output 4-20mA for operating density, reference density, concentration, °Brix, °Plato or other quantities derived from density

With preamplifier:
Vibration frequency modulated to output current and output of the medium temperature by PT100 in 4-wire technology or by 4-20 mA transmitter for evaluation by an external flow or density computer (e.g. series UR 06).

Ex-protectionFlame proof or intrinsically safe

SPECIAL FEATURES

  • high measurement accuracy
  • 2-wire technology
  • special alloy of NiFeCr and 1.4571
  • Measurement independent of flow rate
  • Version without seals
  • material certificates